FDA Enforcement · May 27, 2026

Device recall: Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064),…

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and display… — PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Class II · Ongoing

FDA Enforcement · May 27, 2026

Device recall: Swan-Ganz Pacing Catheter, Models: D200F7;

Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss. — Becton, Dickinson and Company · Class II · Ongoing

FDA Enforcement · May 27, 2026

Device recall: Integris-Allura system; System Code Description (Model Numbers): Integris CV Cesar-Powerpack-Visub-Nicol…

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and display… — PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Class II · Ongoing